Lotte - Big update - Clover detector characterization
2 new clovers - goal is to characterize them.
Using Cs-137 and Eu-152.
Looking into efficiency and temperature dependance
Began with a rough energy calibration using just 2 points in Cs
Then addback efficiency as they are clover detectors. Resulting in a better total efficiency at the cost of lowing the resolution.
This was performed for many different distances ranging from ~10 mm to ~145 mm
See improvement in efficiency with addback, but also see jump in efficiency at 40 mm for clover 147 and ~85 mm for clover 147.
-> Is is due to the step in the hypermet that describes the true shape?
But the jump see in the efficiency is very large and it isn't clear why this would be distance dependent?
One of the clovers has a significant left side tailing compared to the other (clover 146).
What to look into:
A full calibration and a proper calculation of the uncertanties including the source activity error.
Checking the efficiency for the Eu-152 source.
Calculation of the geometrical efficiency
3: Temperature dependance
Measured long measurement with thermocouples on HV supply, digitizer and preamps for both detectors.
Presenting:
Clover 147: 121 keV Eu source
Clover 146: 662 keV Cs source
We couldn't see a relationship between any of the temperatures measured and the centroid of the lines measured.
Absolute temperature variation isn't known but the ohm readout from the thermocouples is measured
Direct comparison between these temperatures and the crystal temperature might not be possible and maybe the temperature change due to the LN2 filling has a greater effect.
Humidity is also important and correlated with temperature but this was not monitored.
So far only one crystal of each detector has been analyzed and this still is to be done for all crystals.