      Short round table updates

      Stella:

      Working with Jana with efficiencies,

      Communication with War War over simulations of IDS

      Meeting Monday with IDS team to discuss IDS tape tests and timings.

      What we want to test, especially in regards to the Ni and Bi beamtimes

      Kayla:

      Energy calibration fitting for Pb castle

      Co-60 -> Gaussian not good 

                 -> Low E tail and high E tail is needed

      the cause of the high E tailing is unknown

      Jack

      Half-life of 220Po from 216Po alpha decay = 25.4 +/- 0.7 mins

      Jana

      Discrepancy in efficiencies assumed to be the change in position of detectors after 1st calibration measurement

      This is now supported by including more calibration measurements and a closer agreement of the efficiencies is found.

      Simone

      Returned from Istros conference.

      Went to present the DSS commissioning and the level structure in In

      Technical request:

      CAEN scintillator lab -> is it possible to use the plastic scintillator with a LaBr to do beta-gamma coincidence and show the students an application of the detectors

      Fran

      Getting everything sorted for registering at CERN as a user

       

       

       

    • 2
      Upcoming experiments / Setup discussion

      No significant updates

    • 3
      Analysis updates

      Lotte - Big update - Clover detector characterization

      2 new clovers - goal is to characterize them.

      Using Cs-137 and Eu-152.

      Looking into efficiency and temperature dependance

      Began with a rough energy calibration using just 2 points in Cs

      Then addback efficiency as they are clover detectors. Resulting in a better total efficiency at the cost of lowing the resolution.

      This was performed for many different distances ranging from ~10 mm to ~145 mm

      See improvement in efficiency with addback, but also see jump in efficiency at 40 mm for clover 147 and ~85 mm for clover 147.

      -> Is is due to the step in the hypermet that describes the true shape?

      But the jump see in the efficiency is very large and it isn't clear why this would be distance dependent?

      One of the clovers has a significant left side tailing compared to the other (clover 146).

      What to look into:

      A full calibration and a proper calculation of the uncertanties including the source activity error.

      Checking the efficiency for the Eu-152 source.

      Calculation of the geometrical efficiency

      3: Temperature dependance

      Measured long measurement with thermocouples on HV supply, digitizer and preamps for both detectors.

      Presenting:

      Clover 147: 121 keV Eu source

      Clover 146: 662 keV Cs source

      We couldn't see a relationship between any of the temperatures measured and the centroid of the lines measured.

      Absolute temperature variation isn't known but the ohm readout from the thermocouples is measured

      Direct comparison between these temperatures and the crystal temperature might not be possible and maybe the temperature change due to the LN2 filling has a greater effect.

      Humidity is also important and correlated with temperature but this was not monitored.

      So far only one crystal of each detector has been analyzed and this still is to be done for all crystals.